PacBio Sequencing

GC3F has provided PacBio sequencing since 2017 and has been a Certified Service Provider since 2021. With the acquisition of PacBio's Revio sequencer, we are now able to offer even more affordable and high-throughput sequencing. We are happy to work on "unconventional" projects and difficult-to-extract species.  


PacBio Revio 

The Revio is PacBio's high-throughput platform, capable of running 16 SMRT Cells per week. Revio data automatically includes 5mC, 5hmC, and 6mA sites (methylation data) for native DNA. We offer both single-use SPRQ and reloadable SPRQ-Nx SMRT Cells. 

Revio SMRT Cell yield depends on the library type:

  • Whole-genome sequencing: 60–120 Gb
  • Amplicons <2 kb: 4–5 million reads
  • Amplicons 2–5 kb: 6–8 million reads
  • 8-plex Kinnex: 50–60 million reads
  • 12-plex Kinnex: 70–80 million reads
  • 16-plex Kinnex: 100–120 million reads

Re-useable SMRT Cells

The new SPRQ-Nx SMRT Cells are used up to three times with washes between libraries. Per PacBio, <0.1% of reads bypass washing and are found in subsequent loadings (0.003-0.09% in GC3F testing). GC3F delivers only Q20+ reads containing your library barcode(s) for reusable cells to further minimize carryover. 

Filtering for only barcode-matching HiFi reads removed all detectable carryover in our test data, but you may still receive reads from another user's library and vice-versa. If this is a concern for your project or you need raw, failed, unbarcoded or unsegmented (Kinnex) reads, we recommend requesting single-use SMRT Cells. 

Revio Submission Requirements

PacBio library prep and sequencing success is highly dependent on the quality of input DNA for library prep. When preparing a WGS library for sequencing, please ensure that your input DNA adheres to the input requirements listed under PacBio WGS Submission Requirements on our prep request page.